Apg MPX Slave Sensor user manual Instrukcja Użytkownika Strona 5

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 48
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 4
Rev. B, 4/13 MPX Series Magnetostrictive Level Sensors
5
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]
Installaon
The MPX should be installed in an area, indoors or outdoors, which meets the
following condions:
The ambient temperature does not exceed -40
o
C to 85
o
C (-40
o
F to
+185
o
F)
Relave humidity up to 100%
Polluon Degree 2
Measurement Category II
Altude 2000 meters or less.
Locate the Probe away from strong magnec elds, such as those pro-
duced by motors, transformers, solenoid valves, etc.
The medium is free from metallic substances and other foreign maer.
No corrosive gases such as NH3, SO2, Cl2, etc.
No excessive vibraon
Ample space for maintenance and inspecon.
NOTE: Assure proper orientaon when installing the oat(s) on the stem
(refer to drawing below). Installing the oat(s) upsidedown can result in
unstable readings.
UP
Taper Markings
XXXX
NC
Przeglądanie stron 4
1 2 3 4 5 6 7 8 9 10 ... 47 48

Komentarze do niniejszej Instrukcji

Brak uwag